If Only I Had a Few Negative X-rays, I’d be All Set

% of readers think this story is Fact. Add your two cents.


In Chapter 16 of Intermediate Physics for Medicine and Biology, Russ Hobbie and I discuss intensity-modulated radiation therapy. The goal is to fire x-rays from many different directions, each direction having a different distribution of intensity, to provide a large dose to the tumor while sparing the surrounding normal tissue.

In classical radiotherapy, the beam was either of uniform fluence across the field, or it was shaped by an attenuating wedge placed in the field. Intensity-modulated radiation therapy (IMRT) is achieved by stepping the collimator leaves during exposure so that the fluence varies from square to square in Fig. 16.45 (Goitein 2008; Khan 2010, Ch. 20)

It was originally hoped that CT [computed tomography] reconstruction techniques could be used to determine the collimator settings at different angles. This does not work because it is impossible to make the filtered radiation field negative, as the CT reconstruction would demand. IMRT with conventional treatment planning improves the dose pattern (Goitein 2008; Yu et al. 2008), providing better sparing of adjacent normal tissue and allowing a boost in dose to the tumor.

What’s up with this talk in IPMB about “negative” radiation? To gain insight, read what Steve Webb wrote in his review article “The Physical Basis of IMRT and Inverse Planning” (British Journal of Radiology, Volume 76, Pages 678-689, 2003).

With the glorious wisdom of hindsight some very early developments—we may call them pre-history—might be considered part of the development of IMRT. The mathematician George Birkhoff showed in 1940 that any drawing could be made up of lines of varying pencil thickness so long as negative pencils were allowed [20]. If we read ‘‘X-rays’’ for ‘‘pencils’’ and ‘‘dose distribution’’ for ‘‘picture’’ the analogy with IMRT is clear. Sadly there are no negative X-rays or uncomplicated tumour control would be 100% guaranteed…

[20]. Birkhoff GD (1940) “On Drawings Composed of Uniform Straight Lines,” Journal de Mathématiques Pures et Appliquées, Volume 19, Pages 221–236.

Oh well,

The best laid schemes o’ mice an’ men 
Gang aft a-gley

If only I had a few of those negative x-rays, I’d be all set.

An example of how a drawing can be made up of lines (in this case, of uniform thickness), based on a figure from “On Drawings Composed of Uniform Straight Lines” by George Birkhoff. An analogous problem is faced when designing a treatment plan using intensity-modulated radiation therapy.


Source: http://hobbieroth.blogspot.com/2019/05/if-only-i-had-few-negative-x-rays-id-be.html


Before It’s News® is a community of individuals who report on what’s going on around them, from all around the world.

Anyone can join.
Anyone can contribute.
Anyone can become informed about their world.

"United We Stand" Click Here To Create Your Personal Citizen Journalist Account Today, Be Sure To Invite Your Friends.

Please Help Support BeforeitsNews by trying our Natural Health Products below!


Order by Phone at 888-809-8385 or online at https://mitocopper.com M - F 9am to 5pm EST

Order by Phone at 866-388-7003 or online at https://www.herbanomic.com M - F 9am to 5pm EST

Order by Phone at 866-388-7003 or online at https://www.herbanomics.com M - F 9am to 5pm EST


Humic & Fulvic Trace Minerals Complex - Nature's most important supplement! Vivid Dreams again!

HNEX HydroNano EXtracellular Water - Improve immune system health and reduce inflammation.

Ultimate Clinical Potency Curcumin - Natural pain relief, reduce inflammation and so much more.

MitoCopper - Bioavailable Copper destroys pathogens and gives you more energy. (See Blood Video)

Oxy Powder - Natural Colon Cleanser!  Cleans out toxic buildup with oxygen!

Nascent Iodine - Promotes detoxification, mental focus and thyroid health.

Smart Meter Cover -  Reduces Smart Meter radiation by 96%! (See Video).

Report abuse

    Comments

    Your Comments
    Question   Razz  Sad   Evil  Exclaim  Smile  Redface  Biggrin  Surprised  Eek   Confused   Cool  LOL   Mad   Twisted  Rolleyes   Wink  Idea  Arrow  Neutral  Cry   Mr. Green

    MOST RECENT
    Load more ...

    SignUp

    Login

    Newsletter

    Email this story
    Email this story

    If you really want to ban this commenter, please write down the reason:

    If you really want to disable all recommended stories, click on OK button. After that, you will be redirect to your options page.